| NO. | Item | Value |
|---|---|---|
| 1 | Oscillations | 11000cpm |
| 2 | Swing | ≥4.5° |
| 3 | Noise | ≤75dB |
| 4 | Sterilization temperature | 135°C |
| 5 | Operation voltage | 9.6V |
| 6 | Battery capacity | 2100mAh |
The standard sterilization temperature should not exceed 135 °C. An improved type can withstand up to 155 °C. Exceeding these limits will cause damage to the motor.
No, the battery is not resistant to high temperatures and cannot be sterilized. It must be safely removed from the handpiece prior to sterilization procedures.
The standard bone drill holds drill bits under 6 mm. The high-torque reaming drill operates at approximately 300 rpm, while the hollow drill (used with K-wire or nails) operates at around 600 rpm.
The cranial bur must be cleaned with a purger immediately post-surgery to prevent blood from drying and sticking. Ensure it is dried thoroughly before storage.
The cranial bur blades are made of medical stainless steel, which is designed to be flexible rather than extremely hard to prevent shattering. As a result, they wear out easily, and single-use is highly recommended to ensure optimal performance.
Always remove the battery from the tool's handpiece after surgery to prevent slow automatic discharge, which shortens overall battery life. Additionally, make sure the battery compartment inside the handpiece is completely dry after sterilizing.